Ranger 7 Variable Range Simulator

Características

• Designed for high resolution
• Used in high precision testing, troubleshooting and calibrating of digital weight indicators
• Switchable offset
• Painted die cast enclosure



Descripción

The Ranger 7 is the simulator of choice for precision testing. Designed for applications that require high resolutions, it features a switchable offset and two Vernier controls, along with switches that go down to 0.01 mV/V.



Especificaciones

Vernier:
Two ranges; one is equal to the sum of the 0.05 mV/V switch and the 0 mV/V-0.5 mV/V switch. The other is the sum of the 0.5 mV/V, 0.1 mV/V and 0.01 mV/V switches

Load Impedance:
Three settings: X0.5= 700 ohms, X1= 350 ohms (one standard load cell),
X4= 87.5 ohms (four load cell platform)

Linearity:
± 0.01% + 1µV

Connectors:
Low Thermal Electronics, Type 2758 precision gold plated binding posts Standard banana jacks or
14 AWG wire

Temperature Coefficient:
5 ppm/°C

Zero Offset:
± 0.0005 mV/V

Output Range:
0 mV/V-3 mV/V in 0.01 mV/V steps

LED Annunciators:
EXC +
EXC –
Offset ON
0.5 mV/V
0.1 mV/V